PEth Concentrated LC-MS Kit - Solvent Kit
- Regulatory Status
- RUO
- Kit size
- 200
- Method
- LC-MS
- Incubation time
- LC-MS run time: 5 minutes
- Specimen / Volumes
- 100 µl Whole blood
- Substrate / isotope
- N/A
This Research Use Only product is intended to assist in the detection of previous ethanol exposure by the quantitative determination of different Phosphatidylethanol (PEth) species in whole blood, using LC-MS. The product is not intended for self-testing, companion diagnostic use, or as a fully automated test.
PLPEth and POPEth are determined from human whole blood by UHPLC with negative ion electrospray LC-MS/MS.
Prior to the LC-MS/MS analysis, a sample clean-up is performed to remove the sample matrix and to spike with the internal standard.
After separation by chromatography on an analytical C-18 column, PLPEth and POPEth are ionized by electrospray ionization (ESI) and detected by LC-MS/MS.
Electrospray ionization is a soft ionization technique where a strong electric field is applied to the liquid passing through the ESI-capillary of the MS-source. The ions are mostly formed in solution before desorption and then transferred into the ion path of the tandem mass spectrometer, which consists of three quadrupoles (two mass selectors connected by a collision cell).
Measurement of the analytes is carried out in MRM (Multiple Reaction Monitoring) mode. In this mode, only selected ions (precursor ions) with a defined mass/charge (m/z) ratio are isolated in the first quadrupole and subsequently transferred into the collision cell, where they are fragmented by impact with an inert gas (argon or nitrogen) at defined voltage settings. Among the fragments generated (known as product ions), only those with a defined m/z ratio can pass the third quadrupole for final detection. In this way, the MRM mode ensures a selective identification and quantification of the target analytes.
Two different concentrated kits, as well as single components, are available to order.
This Kit includes all solvents for HPLC and sample extraction, calibrators and controls.
The Solvent Kit includes all solvents required for sample extraction and liquid chromatography, including the deproteinization solution and mobile phases. Calibrators, controls, consumables, or a chromatographic column are not included.
All listed components can be ordered separately by the order number mentioned below. An HPLC column and a consumables set are also available separately.
| Separate available components |
|---|
|
LC Column PEth 30286448 |
Our comprehensive immunoassay portfolio includes a number of specialty diagnostic immunoassays for endocrinology, immunology and autoimmunity, as well as for diagnosis of multiple infectious diseases. We are pioneers and market leaders in saliva diagnostics, with over 40 years of experience supplying a broad portfolio of luminescence- and ELISA-based tests, including our highly acclaimed HMGB1 and MuSK-Ab ELISAs.
And as experts in laboratory automation, we can support our customers with the protocols for open ELISA platforms.
All products are only available for sale to laboratory professionals and may not be available in all countries. Availability and regulatory status may vary across regions depending on local country-specific registration. Please always read and follow the instructions for use.
All of our assays have been designed and manufactured to meet the highest global regulatory requirements and quality standards. Tecan is certified under ISO 9001:2015, ISO 13485:2016 and is audited by a notified body according to Medical Device Single Audit Program (MDSAP).

As part of the Tecan Group, we have a leading market position in diagnostics and research, with over 40 years of experience in the development, manufacture and supply of enzyme-, radiolabel- and luminescence-based immunoassays.
Our range of high-quality immunoassays is supported by a diverse portfolio of automated solutions, making us the perfect partner for you and your customers.
At Tecan, we are driven to improve people’s lives and health.